Sprecher
Dr.
Eike Feldmeier
(HIT)
Beschreibung
The Heidelberg Ion-Beam Therapy Centre HIT uses the RF-KO slow extraction method to extract the particles from the synchrotron. To improve the spill quality of the extracted beam a new RF-signal was investigated which increases the R-value from 92.5% to 97.5%. The signal is a multiband RF signal broadened with a random BPSK at 3 frequency bands. The new KO-DDS which generates the signal is in medical use since October 2023 and the first experience are presented in this talk.
Hauptautoren
Dr.
Eike Feldmeier
(HIT)
Edgar Christopher Cortes Garcia
(GSI Helmholtzzentrum für Schwerionenforschung GmbH(GSI))
Christian Schömers
(Heidelberg Ion Beam Therapy Centre)
Marcel Hun
(Heidelberg Ion-Beam Therapy Center)
Thomas Haberer
(Heidelberg Ionbeam Therapy Center)
Andreas Peters
(Heidelberger Ionenstrahl-Therapiezentrum (HIT))